受孕期分析仪
型号:: MK-11
价格:请致电:010-57128832,18610462672
品牌: 日本    产品商标: muromachi

Muromachi MK-11 老鼠受孕期分析儀

    
特色:
  • 本儀器專門測RAT陰道表皮細胞的電阻以推測其受孕期。
  • 可測0 - 19.9 Kohm 的電阻值,若電阻值大於3 Kohm 時,表示為proestrous stage。
  • 再現性很好
  • 操作簡單,只要將probe插入老鼠的陰道即可測量,相當方便。

MK-10C規格
Measurement Range0 - 19.9 kohm
Accuracy(+)(-) 1 % (+)(-) 1 digit
DisplayLCD
Battery 4 Rechargeable Battery (Ni-MH)
Battery Life30 hours (When used continuously)
Dimensions90W x 135D x 35mmH 
WeightApprox. 200 g (Battery excluded)

PROBE
 RP-45A Rat Probe
 55mmL x 45mmO.D.
 MP-35 Mouse Probe
 50mmL x 35mmO.D.

 
 

MK-11 Rat Vaginal Impedance Checker has been designed to obtain a precise information on optimum day for mating during estrous cycle in the rats.

The electrical impedance of the epithelial cell layer of vaginal mucosa is measured at the frequency of 1 kHz by insertion of the probe into vagina. In the proestrus stage significantly high impedance is produced compared to that in the other stages of the estrous cycle. 3 kohm of impedance can be considered a standard indicating proestrus stage. Measuring range is 0 - 19.9kohm.

The conventional method known as vaginal smear method takes troublesome procedures such as taking smear, drying, dying and check-up using microscope. It really takes skill as well as time and labor.

Rat Vaginal Impedance Checker is a time & labor saving device on top of high reliability. Further, operation is so simple that even inexperienced personnel can easily use it.

 NOTE: This product has been developed in cooperation with Chugai Pharmaceutical Co., Ltd.

 


P: Proestrus

?: Changes of EIV during estrus cycle in the rats. EIV shows high value at proestrus stage and periodical alteration concurrently with estrus cycle

   M. Koto (1987), Exp. Anim., 36(2), 195-198

 

The Standard System Includes:
MK-11 Main Unit 1
RP-45A Rat Probe 1
CA-100A Cable 1
Rechargeable Battery (Ni-MH) 4

 

REFERENCES:

Bartos, L.(1975). Physiol. Bohemoslov., 24,427.
Bartos, L.(1977). Lab. Anim., 11,53-55
Bartos, L.(1977). Lab. Anim., 11,57-58
M..Koto(1987). Exp. Anim., 36(2),151-156.
M..Koto(1987). Exp. Anim., 36(2),195-198
Teradach,C.(1982). Sci. Tech. Anim. Lab.,7,49-84

 

The continuous use of the MK-11 Rat Vaginal Impedance Checker does not affect pregnancy rate or other reproduction achievements.