Price:
Precision Material Thickness Gauge
For textiles and non-woven textiles according to DIN EN ISO 5084
For fleece according to DIN EN ISO 9073-2
Thickness Gauge to determine the material thickness of all non-woven fabrics. With these instruments the most popular checking methods can be applied for textiles, felt, geomaterials, paper, floor coverings, leather, foils, etc. According to international standards DIN, EN and ISO Common Standards For textiles and non-woven textiles according to DIN EN ISO 5084 (measuring face 20 cm²/pressure 0.1 kPa und 1 kPa) For fleece according to DIN EN ISO 9073-2 (measuring face 25 cm²/pressure 0.1 kPa and 0.5 kPa) For geomaterials according to DIN EN ISO 964-1 (measuring face 25 cm²/pressure 2 kPa and 20 kPa, not 200 kPa) 
Features

- Rigid housing frame
- Separate measuring block, incased against outside external impacts, for highest accuracy and repeatability
- Switch-selectable measuring units mm / inch
- RS 232 data output
- Resolution 0.01 mm - standard, 0.001 mm - optional
- Feeler movement manual by plunger
- Includes Calibration Certificate
- Optional software for data transfer and analysis - Part No. SW-50

Specifications

Resolution | 0.001 mm / 0.0001 inch | Measuring range | 0 - 25 mm / 1 inch | Depth of jaws | 200 mm | Display | LCD 6 digit, 8.5 mm high | Measuring principle | Capacitive | Measuring pressure | < 1.4 N | Feeler shape | a,b,c,d or e | Lifting device | Operating by lifting device - self closing | Data output | RS 232, optical coupling | Power supply | Battery 3 V Lithium, CR 2032 | Temperature range | 10 - 45°C | Weight, net | 1000 g |

Available Models
 Model | Standard | Application | Feeler Ø mm | Feeler cm² | Pressure | D-2000-T | DIN EN ISO 5084 | Textiles | 50.42 | 20 | 0.1 kPa and 1 kPa | D-200-V | DIN EN ISO 9073-2 | Standards Fleece | 56.42 | 25 | 0.1 kPa and 0.5 kPa | D-2000-G | DIN EN ISO 964-1 | Geomaterials | 56.42 | 25 | 2 kPa and 20 kPa | D-2000-NW | DIN EN ISO 53855 | Non-woven textiles | 56.42 35.68 | 25 10 | 0.5 kPa and 1 kPa
5 kPa | D-2000-P | DIN EN ISO 53105 | Paper | 16 | 2 | 100 kPa | D-2000-HY | DIN EN ISO 54540 | Toilet tissue | 56.42 | 25 | 2 kPa | D-2000-L1 | DIN EN ISO 53326 | Leather | 10 | 0.785 | 50 kPa | D-2000-L2 | ISO 2589 | Leather | 9.80 | 0.75 | 49 kPa | D-2000-Gi | | Rubber | 35.68 | 10 | 5 kPa |

Measuring Ranges
 Code R | Higher resolution 0.001 instead of 0.01 mm | Code M | Motorized feeler movement 2.5 mm/sec. operated by foot plug | Code T | Adjustable measuring time 2 to 32 seconds in 2 sec. steps | Code E | Higher measuring range 0... 25 mm | Code V | Adjustable measuring range 0 - 10 mm or 0 - 25 mm | Code P | Additional weights | Code F | Additional feeler shape according to customer specification | DKD-Cal | Calibration report from a DKD authorized laboratory |