Applicable standards
ISO 5627, Tappi T479, DIN 53107
Characteristics
Fully automated measuring cycle, sample is not influenced manually |
|
Big Color Touch Screen |
|
|
Full Statistics, with graphs, average, deviation, min and max value, etc |
3 volume selectable (380, 38 and 19 ml) |
|
|
Selectable measuring ranges: –50.7 to -48.0 kPa & 50.7-29.3 kPa |
Built in weight of 10 kilos |
|
|
Photocell for automatic test when positioning the sample, or manual test |
Measuring area of 10 cm2 |
|
|
RS-232 interface for connection to management and control programs |
Adjustable feet for the equipment levelling |
|
|
USB connection for maintenance works and for connection with USB printer |
CE mark |
|
Test description
The sample is placed below the measuring head. On pushing the start button, the 10 kg weight lowers onto the sample, which helps to hold it tight to the glass plate.
The built-in vacuum pump evacuates the selected tank to a vacuum of -50.7 kPa. Depending on the surface roughness of the sample, atmospheric air is drawn in between the glass plate and the sample.
The time taken for the vacuum to reach -48.0 kPa (res. 29.3 kPa) is displayed in Bekk seconds.
Ten seconds after the test has started a preview is displayed to give an advanced estimate of the result.
Specifications
• Selectable measuring ranges : 50.7-48.0 kPa and 50.7-29.3 kPa |
|
• Easy handling due to touchscreen |
|
• Measuring mode: Underside |
• Selectable volumes: 380 ml (1:1), 38 ml (1:10), 19 ml (1:20) |
• Measuring area: 10 cm2 |
• Parallel printer connection |
• Preview calculation already after 10 sec |
• Measuring accuracy: 0.01 sec |
• Serial data interface RS 232 |
|
|
|
Connections
Electricity: 110-230 V, 50-60 Hz AC |
|
Air: 400-600 kPa |
Dimensions: 27 x 68 x 60 cm (W x D x H) |
Net weight: 23kg Gross weight: 38kg |
|
|
For further information,